BACKGROUND: Type 2 diabetes (T2DM) and cognitive impairment are common long-term chronic conditions affecting older people in hospital. Cognitive impairment can complicate glucose monitoring and lead to diabetes-related emergencies in T2DM. Traditionally, point of care test measurements of capillary blood glucose are conducted in-hospital for T2DM while continuous glucose monitoring (CGM) is not widely used. AIM: To understand the feasibility, acceptability and tolerability of using CGM in older inpatients with T2DM and cognitive impairment. METHODS: 32 older people (mean age=78.7±6.7 years) with comorbid T2DM and cognitive impairment (Abbreviated Mini-Mental Test ≤8/10 and Mini-Addenbrooke's Cognitive Examination ≤22/30) were recruited within a tertiary care hospital in the UK. All participants were naive to CGM and were asked to wear blinded Dexcom G7 sensors for up to 10 days. Participants were asked about feasibility, acceptability and tolerability questions at the point of sensor removal. RESULTS: 29 participants (96%) reported no pain during CGM fitting. All participants (100%) agreed that they did not notice wearing the sensor, and it did not affect their day-to-day hospital activities. All participants (100%) found it 'very easy' or 'easy' to have the sensor fitted and wearing it for 10 days, with 27 participants (90%) finding CGM convenient. 17 participants (57%) reported favourable perceptions of the subcutaneous sensor sensation. CONCLUSION: CGM use in older inpatients with T2DM and cognitive impairment is highly feasible and acceptable for patients. Future studies and trials are now needed to evaluate the clinical use of CGM for glucose monitoring in hospitalised or community-dwelling older individuals with T2DM and cognitive impairment.
Parole chiave
Leggi il testo integrale originale
Testo estratto dalla pubblicazione: sono esclusi immagini, media, supplementi e parti soggette a diritti distinti.
Continuous glucose monitoring in older inpatients with type 2 diabetes and cognitive impairment: an open single-arm feasibility study
Abstract
Abstract
Background
Type 2 diabetes (T2DM) and cognitive impairment are common long-term chronic conditions affecting older people in hospital. Cognitive impairment can complicate glucose monitoring and lead to diabetes-related emergencies in T2DM. Traditionally, point of care test measurements of capillary blood glucose are conducted in-hospital for T2DM while continuous glucose monitoring (CGM) is not widely used.
Aim
To understand the feasibility, acceptability and tolerability of using CGM in older inpatients with T2DM and cognitive impairment.
Methods
32 older people (mean age=78.7±6.7 years) with comorbid T2DM and cognitive impairment (Abbreviated Mini-Mental Test ≤8/10 and Mini-Addenbrooke’s Cognitive Examination ≤22/30) were recruited within a tertiary care hospital in the UK. All participants were naive to CGM and were asked to wear blinded Dexcom G7 sensors for up to 10 days. Participants were asked about feasibility, acceptability and tolerability questions at the point of sensor removal.
Results
29 participants (96%) reported no pain during CGM fitting. All participants (100%) agreed that they did not notice wearing the sensor, and it did not affect their day-to-day hospital activities. All participants (100%) found it ‘very easy’ or ‘easy’ to have the sensor fitted and wearing it for 10 days, with 27 participants (90%) finding CGM convenient. 17 participants (57%) reported favourable perceptions of the subcutaneous sensor sensation.
Conclusion
CGM use in older inpatients with T2DM and cognitive impairment is highly feasible and acceptable for patients. Future studies and trials are now needed to evaluate the clinical use of CGM for glucose monitoring in hospitalised or community-dwelling older individuals with T2DM and cognitive impairment.
Keywords: Hospitalization; Dementia; Inpatients; Diabetes Mellitus, Type 2
STRENGTHS AND LIMITATIONS OF THIS STUDY
Patients and healthcare professionals blinded to CGM readings.
Diverse recruitment across multiple health conditions and socioeconomic backgrounds.
Small sample size and single-centre recruitment limit generalisability.
Inclusion of acutely unwell patients may have affected study recruitment and outcomes.
Introduction
Type 2 diabetes mellitus (T2DM) is a common long-term condition associated with a loss of insulin sensitivity and hyperglycaemia mostly affecting adults, particularly older adults. Individuals aged 65 years and older with T2DM constitute nearly half of the total population of people living with diabetes.1
Chronic complications of diabetes increase with the duration of disease and can lead to disability, lower quality of life and eventually higher mortality. While the most common chronic complications are cardiovascular diseases, gangrene and stroke,2 one of the long-term common complications of diabetes is dementia.3 Diabetes significantly increases dementia risk, particularly with longer diabetes duration or presence of hypoglycaemia.4 This cognitive impairment emerges through complex interactions involving insulin resistance, brain hypometabolism, inflammatory processes, vascular injury and reduced cerebral blood flow and oxidative stress.5
Cognitive impairment can complicate diabetes self-management, including glucose monitoring, further exacerbating the risk of diabetes complications and dementia pathophysiology progression. Advances in technology for better diabetes management such as continuous glucose monitoring (CGM) can contribute to decreasing the number of complications as well as the risk of cognitive decline.2 Recent data show that a holistic and individualised approach is the ideal pathway for diabetes treatment for older adults with a variety of comorbidities with T2DM and other comorbidities.6 CGM is emerging as a valuable technology due to its safety and practicality, particularly for individuals with multimorbidity or geriatric syndromes such as cognitive impairment.7 However, more research is needed to understand the benefits and challenges of CGM in this population with T2DM and cognitive impairment.
Traditionally, point of care test (POCT) measurements of capillary blood glucose are conducted in-hospital for T2DM while CGM is not currently used in the wards since there is no established feasibility of using such devices within the hospital.8 This feasibility study aims to investigate the feasibility, acceptability and tolerability of CGM in older adults with T2DM and cognitive impairment while being inpatients.
Methodology
Study design
A mixed-method open-label single-arm observational feasibility study of CGM among older people with T2DM and cognitive impairment who are in-hospital patients.
Participants: inclusion and exclusion criteria
Eligible participants were older (≥65 years) inpatients with T2DM for at least 6 months and who had been treated with anti-diabetes drugs (either oral, non-insulin injectables or insulin) (ie, not managed with diet and exercise alone).
Participants were also required to have an Abbreviated Mini-Mental Test (AMT) score equal to or less than 8 (out of 10) and/or a Mini-Addenbrooke’s Cognitive Examination (Mini-ACE) UK Version A score ≤21 (out of 30), indicating any level of cognitive impairment, provided they retained decision-making capacity. The AMT is a commonly used, brief cognitive screening tool within hospitals in the UK. The Mini-ACE is a cognitive screening tool sensitive to earlier cognitive changes using two cut-offs (25 or 21 out of 30)9 and has been validated in various settings.
Potential participants were excluded if they did not have mental capacity and/or were on end-of-life care, as confirmed by the clinical care teams. We also excluded people who were current CGM users, were undergoing dialysis treatment and showed evidence of skin irritations on the upper arm or abdomen where the CGM sensor would potentially be fitted. We also excluded people who had other neurological or psychiatric diseases, as well as those with autism, dyslexia or learning difficulties because these conditions could confound cognitive test results.
Setting
The study was conducted in a tertiary care hospital in England (Norfolk and Norwich University Hospitals NHS Foundation Trust) and the CGM sensor was fitted while the consented participants were inpatients (as detailed in the recruitment process section).
Intervention
The Dexcom G7 CGM device is licensed for use in children and adults and available for consumers to directly purchase. The kits comprise auto applicator kits with sensors with transmitter and receivers to be used in the blind mode. We provided blinded receivers to all participants which were paired with their sensors in advance.
The sensors were applied to the back of the upper arm. Interstitial glucose readings were measured continuously, and data were transmitted from the sensor every 5 min. Time in range (TIR) was defined as the time spent when the glucose concentration was between 4 mmol/L and 10 mmol/L in line with international recommendations and default setting.10–12
Recruitment process
One of the study members investigated the initial selection of potential participants via the hospital electronic system (integrated clinical environment) by checking admissions according to the inclusion criteria, except eligibility criteria of AMT and Mini-ACE and approached the potential participants with a consent to contact form. We prioritised geriatric wards and wards where patients were expected to have longer hospital stays.
Potential participants were subsequently given the participant information sheet and allowed 24 hours to consider taking part in the study. After 24 hours, potential participants were revisited and if they agreed to participate, participants provided written consent. No incentives were offered for participation. After consent, participants were assessed for cognitive impairment via the AMT and Mini-ACE. If participants had a score of 8 or lower on the AMT and/or a score of 22 or lower on the Mini-ACE, they were invited to take part in the study. If the cognitive screening scores indicated no cognitive impairment, participants were told that they were not eligible for participation. Figure 1 provides a detailed Consolidated Standards of Reporting Trials (CONSORT) diagram of the recruitment process.
Participants continued to receive their usual hospital care and their usual diabetes management during the study period.
Baseline assessment
A research team member fitted a CGM sensor on the participant’s upper arm and blinded the receiver. Participants’ socio-demographic details (sex, age, ethnicity, marital status, smoking and driving status) were obtained.
Follow-up assessment
The follow-up assessment was conducted on the 10th day after CGM sensor fitting. The 10-day study period was chosen due to the manufacturer’s recommendation to change the sensor after 10 days and considering that some people would not be hospitalised for longer than 10 days. The participant was visited at their hospital bed or if they had been discharged, at their residence (private or care home). If the participants were discharged within 10 days, they were discharged wearing the sensor and taking the receiver, before a research team member removed the sensor on the 10th day and uploaded the data to the Dexcom Clarity system anonymously. The participants were asked CGM feasibility, acceptability and tolerability questions via a questionnaire, including open-ended questions. Finally, participants underwent the same cognitive testing as during the first research appointment and a further Dried blood spot (DBS) sample was collected. Clinical data for all participants was collected via the hospital’s electronic clinical care management system. Figure 2 demonstrates the study design and procedures briefly.
CGM feasibility, acceptability and tolerability survey
Feasibility, acceptability and tolerability were determined via a questionnaire (online supplemental appendix), including one pain scoring question, two dichotomous, seven Likert-style multiple choice questions and two open-ended questions. We encouraged participants to talk about their experiences and additional thoughts about the device with open-ended questions. The questionnaire was modified from a previous CGM feasibility study.13
Statistical analysis plan
All statistical analyses were conducted using RStudio (R V.4.3.2; R Foundation for Statistical Computing, Vienna, Austria). Given the feasibility nature of the study, analyses were primarily descriptive and exploratory. Descriptive statistics were used to summarise feasibility, acceptability and tolerability survey outcomes, participant characteristics, clinical data and glucose-related metrics. Continuous variables were reported as mean with SD or median, while categorical variables were summarised using frequencies and percentages.
Comparisons between CGM-derived glucose metrics and POCT-derived measurements were performed to assess differences between measurement methods. Pre-CGM and post-CGM cognitive scores for AMT and Mini-ACE were compared using paired t-tests. Statistical significance was defined as a two-tailed p value<0.05.
Sample size calculation
The sample size was based on existing feasibility studies within the field,1314 as well as a prior hospital audit of inpatient data meeting the study criteria.
Ethics
This study has been reviewed and approved by Health Research Authority and Health and Care Research Wales on 23 August 2024 (IRAS project ID: 344263, REC reference: 24/LO/0441).
Patient and public involvement and engagement
Patient and public involvement was conducted during the study development process. Two individuals over 65 years with type 1 diabetes and prior CGM experience provided feedback on study design and acceptability through a consultation meeting, which informed the study planning.
Data availability
No additional data are available.
Results
A total of 32 participants were recruited, with the final sample comprising 30 participants. We excluded two participants: one due to more than 50% missing CGM readings during the 10-day study period; and the second because they were not taking anti-diabetic medication.
Socio-demographic characteristics
The mean age was 78.8±6.7 years, with 18 male (60%) and 12 female (40%) participants. 13 participants (43.3%) were living alone, while the remainder were living with a partner, child, grandchildren (n=15, 30%) or in a care home (n=2, 6.6%). All participants were retired. Two-thirds were secondary school graduates, with 6 (20%) having a higher education degree (Bachelor’s or higher) and 5 (16.6%) did not finish high school. We summarised the main characteristics of the participants in table 1.
Table 1Participant characteristics
Participant characteristics Mean Median Min Max
Age (years) ±SD 78.7±6.7 78 67 94
BMI (kg/cm2) ±SD 31.2±10.9 28.3 20.7 74.3
HbA1c (mmol/mol) ±SD 75.1±29.5 67 45 154
Medications (N) 10.5±4.4 11 3 20
Hospitalisations (N) in the last 12 months (excluding current admission) 0.8±1.5 1 0 8
Length of T2DM (years) 14.9±6.4 12.5 0.6 38
BMI, Body mass index; HbA1c, Glycated haemoglobin; T2DM, Type 2 diabetes mellitus.
Only five participants (16.6%) were walking for pleasure or shopping with three out of the five walking less than 1 hour per week. Seven participants (23.3%) were doing light housework such as cooking and cleaning for less than 3 hours/week and eight (26.6%) were doing gardening or do-it-yourself activities. 43.3% were sedentary; however, 10 participants had at least one leg ulcer, lower limb cellulitis or diabetic foot and 2 had leg amputation which affected physical activity levels. The mean frailty score among our participants was 5.83±0.83 (using 1–9 scale, fitness vs vulnerability), which indicates that the majority of them were mildly frail.
Clinical data
Six (30%) were admitted as inpatients due to diabetes-related issues (eg, diabetic ketoacidosis (DKA), hypoglycaemia, hyperosmolar hyperglycaemic state). Another third was admitted due to a fall, either due to hypoglycaemic or other reasons. One was admitted due to cognitive impairment and confusion. 24 (80%) participants had at least four comorbidities at admission.
One participant had a diagnosis of dementia, which was Lewy body dementia, for 1 year. At the second research appointment, 14 participants (46.6%) were still in hospital, including 2 (6.6%) at a different hospital for rehabilitation. 2 participants (6.6%) were discharged to care home or nursing home, while 14 (46.6%) were at their private home at the end of the study.
30% of participants were treated with insulin, while 46.6% were on oral glucose anti-diabetes agents and 30% of the participants were using both insulin and oral agents (table 2). Due to diabetic foot or foot ulcers, 56.6% were regularly using topical agents. Supplements and multivitamins (mostly vitamin D, folic acid and thiamine) were commonly used (73.3%).
Table 2Medication classes
Total number of participants N (%)
Anti-diabetic medications*†
Insulin (solely) 7 (23.3)
Insulin and oral anti-diabetics (combined) 10 (33.3)
Metformin 16 (53.3)
SGLT-2 inhibitors 9 (30)
DPP4 inhibitors 4 (13.3)
Sulphonylurea 3 (10)
GLP-1 receptor agonists 1 (3.3)
Commonly prescribed other medications*†
Statins 20 (66.6)
Anticoagulants/antiplatelets 20 (66.6)
Topical agents 17 (56.6)
Analgesics 15 (50)
Antibiotics 8 (26.6)
*The medications are the regular medications of our participants recorded at admission.
†There was overlap between the groups; for example, some metformin users were also prescribed DPP4 inhibitors and some inpatients receiving statins were also taking anticoagulants.
DPP4, Dipeptidyl peptidase 4; GLP-1, Glucagon-Like Peptide-1; SGLT-2, Sodium-Glucose Cotransporter-2.
Feasibility, acceptability and tolerability outcomes
Feasibility
All participants (100%) found it ‘very easy’ or ‘easy’ to have the sensor fitted and keeping CGM for 10 days.
Acceptability
All participants (100%) reported they did not feel the sensor physically. 29 participants (96%) reported no pain during CGM fitting. All participants (100%) agreed that they were unaware of the sensor once fitted, and it did not affect their day-to-day hospital activities. All participants (100%) reported CGM did not make any difference to their symptoms or experiences of living with diabetes, any other conditions or self-care. 28 (93.3%) felt confident to wear CGM. 27 (90%) participants reported they were likely to recommend CGM to their friends or family.
Tolerability
17 participants (56.6%) reported that positive sensor skin sensation, while the others reported they did not notice the existence of CGM. 27 participants (90%) found CGM convenient. 24 (80%) reported, overall, they liked using CGM.
We summarised some of the feasibility, acceptability and tolerability feedback on CGM experience from our participants with figure 3.
In our open-ended questions, most participants reported that they forgot the sensor’s existence and did not feel anything, and it was easy to keep CGM for 10 days (≥73.3%). Two participants mentioned unsolicited that they would prefer to use CGM compared with POCT in their daily life since they had sore fingers and numbness and found it painful to prick their fingers multiple times a day. Three participants (10%) reported that they needed to remind healthcare staff that they were wearing a CGM sensor, when they were measuring blood pressure via upper arm blood pressure cuffs. One participant (3.3%) reported that the sensor came off on the fourth day, and that they did not wish to have a replacement sensor.
Three participants (10%) said they were happy to contribute to the research, as diabetes and dementia are very serious diseases and research in this field should be encouraged. One participant who had low nocturnal sensor reading commented “It was good to see hypo readings during the night; I was not aware that I am having it frequently. But I knew it decreases sometimes. I will inform my diabetes nurse and GP about it and discuss CGM, my current medications and night-time snacks”.
CGM versus POCT readings
Mean TIR (4–10 mmol/mol) was 36.2%, time above range (TAR ≥10 mmol/mol) was 62.8% and time below range (TBR ≤3.9 mmol/mol) was 1%. 14 (46.6%) participants wore their CGM continuously in hospital, with CGM active for 89.9% of the time (range 86–95%). Mean TIR based on available POCT readings was 40.8%, TAR was 57.2% and TBR 1.8%, showing similar readings as CGM-derived glucose metrics. However, there were individual differences between CGM and POCT, with some individuals showing no TBR (0%) via POCT, while CGM observed results 3% and 8% TBR within the same participants.
Overall, CGM data were very similar and consistent to POCT glucose measurements. One participant (3.3%) had an apparent CGM sensor fault at day 6 after sensor fitting, leading to consistent above range readings compared with POCT. For this participant, we only included CGM data up to day 6.
Cognitive status
The participants were screened with AMT and Mini-ACE to decide their eligibility for this study. Then, all participants completed other cognitive tests after fitting CGM in the baseline assessment. The mean AMT of the participants was 7.2±2.2 and the mean Mini-ACE was 15.6±4.1 (table 3). We also compared AMT and Mini-ACE scores before and after using CGM using paired t-test analyses (n=28). There was no significant change in AMT scores (p=0.28), while Mini-ACE scores increased significantly (p<0.01), indicating cognitive improvement. Results of our study demonstrate how cognitive impairment common in this frail population, despite the absence of a formal dementia diagnosis.
Table 3Cognitive assessment
Cognitive tests Mean Min Max
AMT (baseline assessment) ±SD 7.2±2.2 3 10
AMT (follow-up assessment*) ±SD 7.5±2.1 3 10
Mini-ACE (baseline assessment) ±SD 15.6±4.1 5 22
Mini-ACE (follow-up assessment*) ±SD 18.5±4.9 8 26
*Two participants (6.6%) did not complete the cognitive testing at the follow-up appointment; one participant reported they were feeling unwell due to diabetic foot pain and did not wish to do cognitive tests, while the other participant was transferred from their ward to critical care unit.
AMT, Abbreviated Mini-Mental Test; Mini-ACE, Mini-Addenbrooke’s Cognitive Examination.
Overall, the results showed high feasibility and acceptability among inpatients with T2DM and cognitive impairment.
Discussion
Our study suggests that CGM use is highly feasible and acceptable for older hospitalised individuals with T2DM and cognitive impairment. Specifically, the results of our study show that wearing CGM devices is highly acceptable, feasible and tolerable, even for people with significant cognitive impairment.
Specifically, our findings replicate a previous CGM feasibility study in the same population within the community, that is, not in-hospital.13 Both their and our findings demonstrate high feasibility and acceptability of CGM. In more detail, the feasibility was evident for all our participants, with no exception. Similarly, all acceptability scores were above 90% indicating that people often did not even notice wearing the CGM device after some time. Tolerability scores were slightly lower but still very high (90%) with participants indicating that they found CGM very convenient to use. Some people were encouraged to use CGM after their discharge from hospital. Further, some participants highlighted that they liked the detection of potential hypoglycaemic events, in particular at night, which would allow them to adjust their night-time snacks. Taken together, our findings demonstrated that CGM is highly acceptable, feasible and well tolerated by participants for people with comorbid dementia and cognitive impairment in-hospital.
The in-hospital setting was in itself challenging, with some potential participants finding it overwhelming to take part in our study part despite expressing initial interest (see figure 1). It highlights the challenges of conducting research in inpatient settings. Yet, final feedback from consented participants showed that CGM did not affect day-to-day activities during their hospital stay, further highlighting that CGM is highly feasible within in this setting but that the setting itself was the main challenge. For this reason, larger prospective studies or clinical trials are warranted to evaluate the impact of CGM within this setting and to determine the optimal clinical implementation strategies for CGM use in this vulnerable inpatient group. In this regard, it is also interesting that half of our participants considered using CGM in the future, that is, after discharge from the hospital. This highlights another research gap, in that people within the community could benefit from CGM use. Indeed, retrospective data analyses have suggested that CGM use in T2DM and cognitive impairment in the community can affect overall mortality and hospitalisation rates.1516 One is RELIEF study showing that 2 years use of flash glucose monitoring was associated with 49% reduced hospitalisation due to acute diabetic events mainly DKA.17 Future prospective CGM studies in the community settings in this regard are needed urgently to reduce mortality and hospitalisation in the first place of this vulnerable group.
In terms of the actual CGM readings, to which participants and hospital staff were blinded, the readings showed similar glucose level findings. While the TIR goal for glucose level between 70 and 180 mg/dL (3.9–10 mml/L) is >70% for the general population,10 consensus suggests less stringent thresholds for older people, with a TIR target of >50%.18 Despite this lower TIR goal, the mean TIR in our population was still low. Lifestyle interventions such as regular physical activity, healthy and balanced diet and modest weight loss have been recommended to improve glucose targets, however, our sample was clearly physically inactive, and many had severe diabetes related complications, including amputations, hindering their physical activity levels and increasing their frailty.19 Further, multiple other reasons, including polypharmacy, perioperative procedures and other hospital changes might affect the glucose readings of patients. Finally, clinical guidelines are clearly geared towards the reduction of hypoglycaemia,819 which might explain the low TIR within our participants.
The CGM readings also detected 28 hypoglycaemic events (≤3.9 mmol/L) for nine participants. 11 of the hypoglycaemic events (39.3%) were nocturnal (12:00–06:00). However, low nocturnal sensor glucose readings in our study may not directly reflect true nocturnal hypoglycaemia and should be interpreted in light of potential sensor bias and sensor-clinical discordance, see also.20 Previous studies suggest that CGM may show less accuracy with lower glucose ranges compared with POCT and highly recommend comparison to capillary glucose levels when possible.821–23 It highlights still the challenges using CGM in settings where people might have low readings that might not be ‘real’ or pose an immediate adverse risk within hospital settings.
In terms of the cognitive screening findings, our results showed that a discrepancy between the employed cognitive measures (AMT vs Mini-ACE). The AMT is widely used to assess cognitive status in hospital settings in the UK.24 Although it has practical use on the wards and at hospital admission, there are some limitations regarding language and culture, as well as unreliability of certain questions and reduced sensitivity to detect cognitive changes cross-sectionally and longitudinally.2526 Because of these limitations, we included the Mini-ACE in our cognitive screening, which is commonly used in UK memory clinics for dementia screening and provides a quick cognitive screening read-out, similar to the Montreal Cognitive Assessment (MoCA) (Note the Mini-ACE has been validated against the MoCA). Our results showed that the Mini-ACE is more sensitive and less likely to have ceiling effects than the AMT. We only applied the lower Mini-ACE cut-off (≤21/30), since this score is accepted as almost certain score for dementia or a person with cognitive impairment.9 AMT results of potential participants were higher than Mini-ACE and would have resulted in exclusion of most potential participants. It suggests that the AMT might be not suitable to cognitively screen in hospital, as it only detects severe cognitive deficits, for example, delirium, while potentially missing patients with mild-to-moderate cognitive deficits. There is a dearth of research on how cognitive impairment levels impact glucose management in older T2DM people. Our results show that despite all participants having mental capacity to consent to the research, their cognitive performance indicated moderate cognitive impairment, further complicating diabetes management.
Interestingly, the cognitive scores for the AMT did not change from baseline to follow-up assessment (10 days later), whereas those for the Mini-ACE did. The Mini-ACE scores at the follow-up assessment were significantly higher than the baseline assessment scores. Since short follow-up duration and use of masked CGM are unlikely to provide meaningful cognitive changes, this might have been due to patients having recovered from their conditions, they were admitted for in the first place. It also highlights that cognition levels can change during hospital stays. In this regard, it was also telling that the AMT scores did not change from baseline to follow-up, further underlining that the AMT has limited sensitivity to measure mild-to-moderate cognitive impairment within hospital cross-sectionally and longitudinally. Future research on cognitive screening in hospital settings is needed to determine whether better tools than the AMT can be applied, as they might improve patient care, including glucose management.
Despite the novelty of our research with promising findings, there are some limitations to our study. Overall, this is a small pilot study in one hospital, which requires replication in a larger sample and different hospital settings. Further, participants were admitted to the hospital due to different health conditions including emergencies. Therefore, they were not in their best health, and this might have affected the cognitive and CGM results. We ran our recruitment search in particular wards due to our inclusion criteria (older age), and having longer hospital stays. This selection may have caused a potential bias in the recruitment process; for example, we recruited five participants from a general surgery ward which potentially resulted in lower physical activity scores in our study population since some participants were awaiting lower limb amputations. Despite this limitation, our findings were highly consistent regardless of health condition or ward recruited from. Finally, in our study, we used masked CGM, and it is, therefore, not clear whether CGM use can actually improve glucose management in hospital, compared with traditional POCT, which requires future RCTs.
Future research should be considered with larger inpatient populations in the community to evaluate whether CGM use can improve glucose management and clinical outcomes compared with standard care, while also assessing its cost-effectiveness in this setting. Additionally, exploring optimal sensor placement and training for healthcare professionals may enhance usability and adherence in hospital settings. Finally, it would be important to investigate whether CGM after hospital discharge, that is, in the community, would benefit people with comorbid diabetes and cognitive impairment, since this is such a vulnerable group for hypoglycaemic events and future hospital admissions.27
In conclusion, CGM is highly feasible and acceptable for older hospitalised patients living with T2DM and cognitive impairment. CGM technology offers an advantage by capturing events such as asymptomatic nocturnal hypoglycaemia. Further investigation is required to understand in more detail the benefits and challenges of CGM in inpatient care for those with T2DM and cognitive impairment.
Supplementary material
Acknowledgements
We would like to thank the participants in this study for devoting their time for this research. The views expressed are those of the author(s) and not necessarily those of the NIHR or the Department for Health and Social Care.
The views expressed are those of the author(s) and not necessarily those of the NIHR or the Department for Health and Social Care.
Footnotes
Funding: This study is supported by the NIHR Research Capability Funding through the Norfolk and Norwich University Hospitals NHS Foundation Trust and Norfolk Diabetes. Dexcom, Inc (San Diego, California, USA) donated half of the sensor kits to our research.
Prepublication history and additional supplemental material for this paper are available online. To view these files, please visit the journal online (https://doi.org/10.1136/bmjopen-2026-118196).
Provenance and peer review: Not commissioned; externally peer reviewed.
Patient consent for publication: Consent obtained directly from patient(s).
Ethics approval: This study has been reviewed and approved by Health Research Authority (HRA) and Health and Care Research Wales (HCRW) on 23 August 2024 (IRAS project ID: 344263, REC reference: 24/LO/0441). Participants gave informed consent to participate in the study before taking part.
Data availability free text: Not applicable. No datasets were generated.
Patient and public involvement: Patients and/or the public were involved in the design, or conduct, or reporting, or dissemination plans of this research. Refer to the Methods section for further details.
Data availability statement
Data sharing not applicable as no datasets generated and/or analysed for this study.
References
1.Bellary S, Kyrou I, Brown JE, et al. Type 2 diabetes mellitus in older adults: clinical considerations and management. Nat Rev Endocrinol. 2021;17:534–48. doi: 10.1038/s41574-021-00512-2.
2.Deshpande AD, Harris-Hayes M, Schootman M. Epidemiology of diabetes and diabetes-related complications. Phys Ther. 2008;88:1254–64. doi: 10.2522/ptj.20080020.
3.Cholerton B, Baker LD, Montine TJ, et al. Type 2 Diabetes, Cognition, and Dementia in Older Adults: Toward a Precision Health Approach. Diabetes Spectr. 2016;29:210–9. doi: 10.2337/ds16-0041.
4.Cao F, Yang F, Li J, et al. The relationship between diabetes and the dementia risk: a meta-analysis. Diabetol Metab Syndr. 2024;16:101. doi: 10.1186/s13098-024-01346-4.
5.Kan W, Qu M, Wang Y, et al. A review of type 2 diabetes mellitus and cognitive impairment. Front Endocrinol (Lausanne) 2025;16:1624472. doi: 10.3389/fendo.2025.1624472.
6.Yun JS, Kim K, Ahn YB, et al. Holistic and Personalized Strategies for Managing in Elderly Type 2 Diabetes Patients. Diabetes Metab J. 2024;48:531–45. doi: 10.4093/dmj.2024.0310.
7.Ergin BD, Gadsby-Davis K, Mattishent K, et al. Continuous glucose monitoring in comorbid dementia and diabetes: the evidence so far. SAGE Publications Inc; 2024.
8.Avari P, Lumb A, Flanagan D, et al. Continuous Glucose Monitoring Within Hospital: A Scoping Review and Summary of Guidelines From the Joint British Diabetes Societies for Inpatient Care. J Diabetes Sci Technol. 2023;17:611–24. doi: 10.1177/19322968221137338.
9.Hsieh S, McGrory S, Leslie F, et al. The Mini-Addenbrooke’s Cognitive Examination: a new assessment tool for dementia. Dement Geriatr Cogn Disord. 2015;39:1–11. doi: 10.1159/000366040.
10.Battelino T, Danne T, Bergenstal RM, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42:1593–603. doi: 10.2337/dci19-0028.
11.Danne T, Nimri R, Battelino T, et al. International Consensus on Use of Continuous Glucose Monitoring. Diabetes Care. 2017;40:1631–40. doi: 10.2337/dc17-1600.
12.Wilmot EG, Lumb A, Hammond P, et al. Time in range: A best practice guide for UK diabetes healthcare professionals in the context of the COVID-19 global pandemic. Diabet Med. 2021;38:e14433. doi: 10.1111/dme.14433.
13.Mattishent K, Lane K, Salter C, et al. Continuous glucose monitoring in older people with diabetes and memory problems: a mixed-methods feasibility study in the UK. BMJ Open. 2019;9:e032037. doi: 10.1136/bmjopen-2019-032037.
14.Welford AE, Ridgeway J, Gillies C, et al. Continuous Glucose Monitoring in the Management of Medication in Care Home Residents with Type 2 Diabetes (eDMED): A Protocol for a Feasibility Study. Diabetes Ther. 2025;16:1511–24. doi: 10.1007/s13300-025-01740-9.
15.Kotecha P, Smith SM, Donahoo WT, et al. Continuous Glucose Monitoring in Insulin-Treated Older Adults With Diabetes and Alzheimer Disease and Related Dementias.
JAMA Netw Open
. 2025;8:e2541939. doi: 10.1001/jamanetworkopen.2025.41939.
16.Reaven PD, Macwan S, Newell M, et al. Initiation of Continuous Glucose Monitoring and Mortality in Type 2 Diabetes. Diabetes Technol Ther. 2025;27:778–89. doi: 10.1089/dia.2025.0227.
17.Riveline J-P, Roussel R, Vicaut E, et al. Reduced Rate of Acute Diabetes Events with Flash Glucose Monitoring Is Sustained for 2 Years After Initiation: Extended Outcomes from the RELIEF Study. Diabetes Technol Ther. 2022;24:611–8. doi: 10.1089/dia.2022.0085.
18.Toschi E, O’Neal D, Munshi M, et al. Glucose Targets Using Continuous Glucose Monitoring Metrics in Older Adults With Diabetes: Are We There Yet? J Diabetes Sci Technol. 2024;18:808–18. doi: 10.1177/19322968241247568.
19.ElSayed NA, Aleppo G, Bannuru RR. 13. Older Adults: Standards of Care in Diabetes-2024. Diabetes Care. 2024;47:S244–57. doi: 10.2337/dc24-S013.
20.Bay C, Kristensen PL, Pedersen-Bjergaard U, et al. Nocturnal continuous glucose monitoring: accuracy and reliability of hypoglycemia detection in patients with type 1 diabetes at high risk of severe hypoglycemia. Diabetes Technol Ther. 2013;15:371–7. doi: 10.1089/dia.2013.0004.
21.Vu L, Kefayati S, Idé T, et al. Predicting Nocturnal Hypoglycemia from Continuous Glucose Monitoring Data with Extended Prediction Horizon. AMIA Annu Symp Proc. 2019;2019:874–82.
22.Garg SK, Kipnes M, Castorino K, et al. Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes. Diabetes Technol Ther. 2022;24:373–80. doi: 10.1089/dia.2022.0011.
23.Gamarra E, Careddu G, Fazi A, et al. Calibrating Dexcom G7 improves its performance in the context of repetitive recreational scuba diving in people with type 1 diabetes. BMJ Open Sport Exerc Med. 2025;11:e002809. doi: 10.1136/bmjsem-2025-002809.
24.Piotrowicz K, Romanik W, Skalska A, et al. The comparison of the 1972 Hodkinson’s Abbreviated Mental Test Score (AMTS) and its variants in screening for cognitive impairment. Aging Clin Exp Res. 2019;31:561–6. doi: 10.1007/s40520-018-1009-7.
25.Emery A, Wells J, Klaus SP, et al. Underestimation of Cognitive Impairment in Older Inpatients by the Abbreviated Mental Test Score versus the Montreal Cognitive Assessment: Cross-Sectional Observational Study. Dement Geriatr Cogn Dis Extra. 2020;10:205–15. doi: 10.1159/000509357.
26.Peters KA, Howe TJ, Rossiter D, et al. The Abbreviated Mental Test Score; Is There a Need for a Contemporaneous Update? Geriatr Orthop Surg Rehabil. 2021;12:21514593211001047. doi: 10.1177/21514593211001047.
27.Gadsby-Davis K, Garner N, Ergin BD, et al. The Impact of Comorbid Dementia and Diabetes Mellitus on Hospital Patients’ Outcomes: A Systematic Review and Meta-analysis. Diabetes Ther. 2025;16:103–20. doi: 10.1007/s13300-024-01672-w.
Come citare questo studio
Busra Donat Ergin, Katharina Mattishent, Anne Marie Minihane, Richard Ian Gregory Holt, Helen Murphy, Ketan Dhatariya, Michael Hornberger. (2026). Continuous glucose monitoring in older inpatients with type 2 diabetes and cognitive impairment: an open single-arm feasibility study.. BMJ open. DOI: 10.1136/bmjopen-2026-118196